Hanna Instruments, Inc. | United States
Customer Support: +1 (401) 555 7463

pH and EC Transmitter with Isolated Output - 0 - 1 V

$0.00

In stock

The HI98143 is part of a family of blind transmitter that can be used in water treatment and hydroponic applications. These meters have inputs for both pH and EC probes. The pH probe input uses a common BNC connector while EC has a direct wire connection for use with Hanna’s EC probes that incorporate an NTC sensor for temperature compensation.

The HI98143 is available in four models based on the analog output signal required.

  • Blind transmitter for pH and EC (up to 10 mS/cm)
  • Choice of analog output in voltage or current
  • Powered or loop powered analog outputs

Direct Connection

Direct connection of the probes to the transmitter assures a positive electrical connection with no signal loss. This transmitter is ideal for remote process control applications.

Output Signals

Four models are available, transmitting a 0-1 V, 0-4 V or 4-20 mA signal. The output signals are proportional to the input signals but independent of changes in load or cable capacitance.

Automatic Temperature Compensation

Compensation for the effects of temperature for EC measurements is performed by the transmitters’ Automatic Temperature Compensation circuitry.

Connectivity

The transmitter can be connected to any pH or conductivity controller, recorder, PC or any data monitoring device that accepts 0 to 1 V, 0 to 4 V or 4 to 20 mA input.

Simple Design

Probes connectors, power supply, contacts, and calibration trimmers are easy to reach directly on the front panel

Versatile Use

For in-line applications, the recommended pH electrode is the HI1001 while the conductivity probe is the HI3001. HI1001 is a combined double junction pH electrode with BNC and 3m cable; HI3001 is a 4-ring conductivity probe with platinum sensors and an int

SKU HI98143-web
Name pH and EC Transmitter with Isolated Output
pH Range 0 to 14
pH Accuracy ±0.5% f.s.
pH Calibration Manual, 2 point, through trimmers
EC Range 0 to 10 mS/cm
EC Accuracy ±2% f.s.
EC Calibration 0 (offset) and 5 mS/cm (slope) trimmers
EC/TDS Temperature Compensation Automatic from 0 to 60 ºC
Temperature Correction Coefficient ß=2%/ ºC
Cell Type 4-ring EC probes
EC/TDS Probe Four (ring) pole probe with built in NTC sensor
Analog Output HI98143-01 0-1 V
HI98143-04 0-4 V
HI98143-20 4-20 mA (passive)
HI98143-22 4-20 mA (active)
Power Supply 12-24 Vdc
Environment 0 to 50 °C (32 to 122 °F);
max 95% RH non-condensing
IP Rating IP54
Dimensions 160 x 105 x 31mm (6.3 x 4.1 x 1.2”)
Weight 280 g (9.9 oz.)

pH and EC Transmitter with Isolated Output

Our Mission

To empower customers to achieve quality by supplying intuitive, accurate, and reliable analytical instruments with exceptional customer service and value.

We take pride in every product we build. From an original idea, to a completed product ready for testing. We oversee every aspect of the manufacturing process. It is this level of attention to detail that sets us apart.

Hanna Instruments Logo
Copyright © 1996-2023 - Hanna Instruments, Inc.
US Flag