HI96714-11 is a high quality set of calibration standards that allows users to validate and calibrate the HI96714 Cyanide Portable Photometer. The HI96714-11 includes one sealed “A” cuvette for zeroing the meter and one “B” cuvette that is used to validate or calibrate. The standard value of the “B” cuvette is 0.100 ±0.005 mg/L (ppm). An acceptable reading when validating the HI96714 with the standards is 0.100 ±0.010 mg/L @ 25℃. The calibration set is produced in our state-of-the-art manufacturing facility using reagent grade chemicals in a temperature controlled environment.
Supplied with Certificate of Analysis
- Product code
 - Lot number
 - Expiration date
 - Standard value @ 25℃
 - Reference meter NIST traceable
 
Provided storage containers
- Light tight
 - Protects from accidental breakage
 
Validation and calibration of the cyanide range for the HI96714 portable photometer is quick and easy with the HI96714-11 CAL Check™ Standards.